ARAI/TA(4G)/MODI/MEXA 554JA/2004-25                                                                    Dated 11 Oct 2004

 

CERTIFICATE

FOR COMPLIANCE TO THE CENTRAL MOTOR VEHICLES RULES

 

M/s Modi Measurement Systems Pvt. Ltd.

219, Hill View Industrial Estate

Amrutnagar Road

Off. L.B.S.Marg

Ghatkopar (W)

Mumbai 400 086

 

Exhaust Gas Analyzer Model MEXA 554JA from M/s Modi Measurement Systems Pvt. Ltd., Mumbai, was examined and tested as per the revised test procedure approved by Ministry of Road Transport and Highways, Government of India, New Delhi, and specified in MoRTH/CMVR/TAP-115/116, Issue No.2, Part-VIII for 4-gas analyzer.

 

Based on the above, it is certified that the 4-Gas Analyzer Model MEXA 554JA manufactured by M/s Horiba Ltd., Japan, and marketed by M/s Modi Measurement Systems Pvt. Ltd., Mumbai, meets the requirements of the provisions of Rule 116(3) of the Central Motor Vehicles Rules 1989, as amended up-to-date, for the 4-Gas Analyzer.

 

The 4-Gas Analyzer Model MEXA 554JA is suitable for measurement of idling CO and HC of in-use vehicles fitted with SI engines as per CMVR 115(2), subject to following conditions :

 

         i.      The 4-Gas Analyzer Model MEXA 554JA shall comply with the requirements as per Clause 9.5(e) of the test procedure within six months from the date of issue of this certificate or before 31st March 2005, whichever is earlier.

       ii.      The supplier shall abide by the Code of Practice attached herewith as Annexure-A.  The supplier has already submitted an affidavit to this effect.

      iii.      The supplier shall enter into AMC agreement on annual basis with the authorized PUC Test Centres to whom this model is supplied.  The AMC terms shall be as per the Code of Practice referred in (ii) above.

 

This Type Approval certificate is valid for the period of 5 years from the date of its issue or any amendment in the test procedure, whichever is earlier.

 

The first Conformity of Production testing will be due on or before 10th Oct 2006 or production of 100 units from 11th Oct 2004, whichever is earlier.

 

Authorised Signatory

 

 

 

 

A. B. Komawar                                                                                                             B. Bhanot

Dy. Director                                                                                                                  Director

 

Ref : i)      Type Approval Test Report No.AED/TR/SPR063/2004-05/Modi-Mumbai/2004-123 Dated 11 Oct 2004


 

ARAI/TA(4G)/MODI/MEXA 554JA/2004-25                                                                    Dated 11 Oct 2004

 

ANNEXURE  - A

 

CODE OF PRACTICE FOR PUC EQUIPMENT MANUFACTURER / SUPPLIER

 

 

1.                   PUC equipment manufacturer / supplier should include the description of the test procedure described in Part I or Part II, whichever is applicable, amended from time to time of the document MOST/CMVR/TAP 115/116 in the user’s manual of the PUC equipment.  PUC equipment manufacturer / supplier shall keep PUC operator informed about the changes in the test procedure in future.

 

2.                   PUC equipment manufacturer/ supplier shall supply copy of type- approval certificate with date of validity along with the PUC equipment.

 

3.                   The validity of the type approval certificate of the PUC equipment shall be 5 years, after the expiry of which the PUC equipment manufacturer/ supplier shall get it re-validated from the test agency.

 

4.                   PUC equipment manufacturer / supplier shall provide the status of production/ supply of PUC equipment at a regular interval of 1 year to the test agency from where the equipment has been certified.

 

5.                   PUC equipment manufacturer / supplier shall submit the equipment for COP as per procedure mentioned above.

 

6.                   PUC equipment manufacturer / supplier shall enter into AMC on annual basis with the authorised PUC test agency based on charges as approved and finalized in EECG meeting as per Annexure-4. This AMC contract shall include 3 visits and equipment calibration as per field calibration procedure given in Annexure-1.  PUC equipment manufacturer/ supplier shall provide calibration certificate as per format given in the Annexure-2.

 

7.                   PUC equipment manufacturer / supplier shall train minimum 3 operators of PUC test agency and shall provide training certificate as per format given in Annexure 3. 

 

 

***********


 

ARAI/TA(4G)/MODI/MEXA 554JA/2004-25                                                                    Dated 11 Oct 2004

 

ANNEXURE – 1 TO ANNEXURE – A

 

FIELD CALIBRATION PROCEDURE FOR TESTING OF

 

GAS ANALYSERS

 

 

1.0              INTRODUCTION

 

This procedure has to be carried out on gas analysers after they are commissioned in the field and for the subsequent calibration.

 

2.0              TESTING

 

The test procedure for gas analysers is as follows:

 

i)                    Check that the power supply is as per specifications of the manufacturer and electrical earthing is proper.

 

ii)                   Check that all the accessories as per manufacturer are available and are functioning properly.

 

iii)                 Check the span and zero calibration using sample gas of suitable value for CO as well as HC.

 

iv)                 Check the electrical calibration.

 

v)                  Check that the sampling system is leak proof.

 

vi)                 The printer is working correctly and the print out details are correct.

 

vii)               Checking of 1 no. of vehicle for idling emission measurement using this analyzer.

 

 

************

 

 

 

 

 


 

ARAI/TA(4G)/MODI/MEXA 554JA/2004-25                                                                    Dated 11 Oct 2004

 

 

ANNEXURE – 1 TO ANNEXURE – A (contd…..)

 

FIELD CALIBRATION PROCEDURE FOR TESTING OF

 

SMOKE METERS

 

1.0  INTRODUCTION

 

This procedure has to be carried out on meters after they are commissioned in the field and for the subsequent calibration. 

 

2.0  TESTING

 

The test procedure for smoke meters is as follows:

 

i)                   After the warm-up of the meter, the calibration of the meter has to be checked at zero and mid-scale point with the neutral density filter available.  The value must lie within 0.1 m-1.

 

ii)                  The meter shall have the standard accessories as specified by the manufacturer.  It shall be checked that the sample hose, internal pipes etc are not deteriorated or damaged to ensure that there is no leakage.

 

iii)                The functionality of oil temperature and RPM sensor.

 

i)                   The heating system for the optical chamber is functioning.

 

ii)                  The purge air system is working correctly.

 

iii)                Visual displays are functioning correctly.

 

iv)                The printer is working correctly and the printout details are correct.

 

v)                 The instrument casing is proper and has proper electrical earthing.

 

vi)                Free acceleration test is carried out using a vehicle and the print out details are checked.

 

 

 

************

 

 


 

ARAI/TA(4G)/MODI/MEXA 554JA/2004-25                                                                    Dated 11 Oct 2004

 

ANNEXURE – 2 TO  ANNEXURE- A

 

CALIBRATION CERTIFICATE FORMAT

 

 

1.0

Component:

PUC equipment model:

Sr. No.

 

2.0

PUC Center

Registration No.:

 

3.0

Objective of the test: To carry out Physical check and calibration of gas                                        Analyser / Smoke meters as per the test procedure specified in Annexure 1 of  CMVR / TAP 115-116 Part-8.

                                      

4.0

4.1

 

4.2

Detailed Observations

Checking of supply/ earthing

 

Checking of                     Details of accessories checked.

accessories:

4.3

Span Calibration

 

     1.   Details of span gas concentration__________%

     2.   Calibration gas cylinder No.: __________

     3.   Calibration gas cylinder make:___________

     4.   Calibration gas validity date:___________

 

OR

 

    5.    Details of Natural Density filters used for mid point calibration

 

4.4

Electrical Calibration       OK/ Not OK

 

 

4.5

Leak test:

Passed/ Failed

5.0

One no of petrol / diesel vehicle checked for idling Emission / Free acceleration, measurement

6.0

Conclusion:

 

 

7.0

Next Calibration Due Date:

 

 

 

Signature & Seal of manufacturer/ Supplier

 

 

 

 


 

ARAI/TA(4G)/MODI/MEXA 554JA/2004-25                                                                    Dated 11 Oct 2004

 

ANNEXURE – 3 TO ANNEXURE - A

 

PUC

Operator

Photograph

 
TRAINING CERTIFICATE

 

 

 

 

 

 

 

It is to certify that Mr. / Mrs. __________________________________________   has attended the training and knows all required operation of the smoke meter / Gas Analyser model ____________________________ to perform PUC tests.

 

Training is given in the following areas:

Understanding of procedure for testing of Idling emission/ free acceleration smoke as CMVR/ TAP/ 115/116 procedure.

 

1.                  Normal thermal condition of the vehicle

2.                  Actual testing procedure

3.                  Procedural understanding of issue of PUC certificate

 

 
1. Vehicle testing mode
2. Zero Calibration
3. Span calibration
4. Electronic calibration
5. Leak test

Maintenance

1. Replacement of filters
2. General maintenance

 

  Authorized Signature & seal of manufacturer / supplier

 

*************

 

 


 

ARAI/TA(4G)/MODI/MEXA 554JA/2004-25                                                                    Dated 11 Oct 2004

 

ANNEXURE-4 TO ANNEXURE-A

 

AMC CHARGES

 

 

 

Sr. No.

Details

4 Gas Analyser

2 Gas Analyser

Smoke Meter

1.

Maintenance done at centralized station of PUC equipment manufacturer

Rs.5,700/- + Rs.750/- per calibration

Rs.3,750/- + Rs.750/- per calibration

Rs.4,500/-

2.

Maintenance done at PUC center’s place

Rs.9,500/- + Rs.1,500/- per calibration

Rs.6,250/- + Rs.1,500/- per calibration

Rs.7,500/-

 

Note:

1.      For 2/4-gas anlyser, PUC equipment manufacturer shall provide the calibration gases required for calibration.

2.      For smoke meter, the AMC charges include calibration with neutral density filter (NDF) by PUC equipment manufacturer.

 

 

 

***************